Item
1. Type of observation
image
North America specification (NTSC)
CCD
3700,01,08
Display
2 steps
size
Medium
↓
Full-height
1. White balance
The user can select the endoscopic image display size from the
followings depending on the scope (CCD) applied.
3704
37071
1 step
2 steps
Full-high
Small
only
↓
Medium
Note 1)
Change the display size on the keyboard or with the scope
switch or RS-232C communication.
Note 2)
Standard setting is as follows.
• 3700/01/08:
• 3704:
• 37071:
1. Specifications
• When replacing the scope with a different type,adjust the
white balance using the supplied white cap.
• A memory function is provided to hold the previous white
balance data even if the power switch is turned off.
• Calibrate the white balance automatically if the scope
with the scope ID function (EVIS EXERA 160 series) is
connected to the CV-160.
Note) It is necessary to calibrate manually again if the light
source changed.
2. Setting method
• Select the type of light source(xenon=CLV/halogen=CLE)
on the system setup screen.
• Depress the white balance switch on the front panel.(It is
not necessary for scopes with the scope ID function.)
3. Switch depression detection time
• Approx. 1 second
4. Time required by correction operation
• Approx. 1 second
1-3
SPECIFICATIONS CV-160
Specification
Europe specification (PAL)
3700,01,08
3704
3 steps
2 steps
Medium
Semi
↓
full-height
Semi
↓
full-height
Full-height
↓
Full-height
NTSC = Medium
PAL = Medium
NTSC = Full-height
PAL = Semi full-height
NTSC = Small
PAL = Small
37071
2 steps
Small
↓
Medium
Need help?